Skip to main content

Table 2 Analysis of the accuracy and of the agreement of the distances estimated by FeetMe® and the rater along the 30-m and 10-m tracks

From: Validation of the six-minute walking distance measured by FeetMe® insoles

Method

n

Track length [m]

Bias [m]

95% CI of differences [m]

R2

Pearson correlation

ICC

95% CI of ICC

MAE [m]

t-test

p-value*

FeetMe®

32

30

-8.30

[-31.01–14.41]

0.96

0.98

0.97

[0.90–0.99]

9.75

0.002

Rater

33

-16.24

[32.47–0.01]

0.99

0.99

0.95

[0.12–0.99]

16.24

FeetMe®

32

10

-9.00

[-36.93–18.93]

0.95

0.97

0.96

[0.86–0.98]

12.86

< 0.001

Rater

33

-37.73

[-79.57–4.11]

0.86

0.93

0.76

[0–0.93]

38.88

  1. Abbreviations: ICC intraclass correlation coefficient, CI confidence interval, MAE mean absolute error, R2 coefficient of determination
  2. *t-test p-values were obtained for comparisons of the absolute errors between FeetMe® estimates and rater estimates